A connector may pass continuity and still show unstable milliohm-level resistance. At this range, cable, relay and probe contact can be as large as the feature being checked. A useful fixture therefore separates current delivery from voltage sensing and also controls how the product is supported and contacted.
Put the force and sense contacts close to the product interface
Route the test current through one probe pair and measure voltage through a separate high-impedance pair. Place the sense points inside the force points where the connector geometry allows, keep the paths short and shielded, and document exactly which plating surface is included. This keeps fixture leads and most probe-contact resistance out of the calculated result.
Make mechanical contact repeatable without damaging the terminal
Support the housing near the test load, use guided spring probes with verified working stroke and prevent side loading. Probe-tip material, spring force and cleaning interval must suit the terminal plating. A positive nest location, closed-position sensor and golden sample help distinguish a worn probe from a real product change.
Define the measurement cycle and acceptance evidence
Use the current level, stabilization time and limit required by the applicable product specification. Offset compensation or polarity reversal can reduce thermal EMF at very low voltage. Record fixture ID, channel, ambient or part temperature, golden-sample result and measured value, then confirm the method with a known resistor and representative good and bad parts.